Tyvaso Inhalation System Patient a) Starter Kit with TYVASO (treprostinil) Inhalation Solution. Treprostinil 1.74 mg/2.9 mL (0.6 mg/mL) (NDC 66302-206-01) b) TD-300/A Replacement Device Kit Material Number RTP3099 RX Only. Manufactured by United Therapeutics Corporation Research Triangle Park, NC 27709
Recall Number: D-1188-2018
Firm: United Therapeutics Corp.
Classification: Class III
Status: Terminated
Report Date: 20180919
Initiation Date: 20180814
Termination Date: 20191030
Reason for Recall
Defective Delivery System: Water ingress through the lower water cup sensor of the device.
Product Description
Tyvaso Inhalation System Patient a) Starter Kit with TYVASO (treprostinil) Inhalation Solution. Treprostinil 1.74 mg/2.9 mL (0.6 mg/mL) (NDC 66302-206-01) b) TD-300/A Replacement Device Kit Material Number RTP3099 RX Only. Manufactured by United Therapeutics Corporation Research Triangle Park, NC 27709